Back to CVE catalog
CVE-2017-9657

Partial re-association of Philips IntelliVue MX40 monitor

Philips IntelliVue MX40 Patient Worn Monitor

At a glance

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Source description · CVE List / CNA

System impact

Attack complexityLow
ConfidentialityNone
IntegrityNone
AvailabilityHigh

CVSS summary v3.0

6.5Medium
Published:
Updated:
EPSS:
0.769%
Exploited in attacks:
Not confirmed
Ransomware campaigns:
Not confirmed
Attack vector:
Adjacent

Affected software

Vendor, product and affected version ranges.

Vendor, product and affected versions
Vendor:PhilipsIntelliVue MX40 Patient Worn MonitorAffected versions:
IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18Show full version range for IntelliVue MX40 Patient Worn Monitor
Vendor:philipsintellivue mx40 firmwareAffected versions:
< b.06.18Show full version range for intellivue mx40 firmware
  • CPE applicability can depend on other products or environments. Consult the full source configuration.

1–2 of 2 · Page 1 of 1

CVSS scores and metrics2
[email protected] · NVD (Primary)Primary6.5 / 10MediumCVSS 3.0Details
Attack complexity
Low
Confidentiality
None
Integrity
None
Availability
High
Attack vector
Adjacent
Privileges required
Not required
User interaction
Not required
Scope
Unchanged

Source vector

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
[email protected] · NVD (Primary)3.3 / 10LowCVSS 2.0Details
Attack complexity
Low
Confidentiality
None
Integrity
None
Availability
Partial
Attack vector
Adjacent
Authentication
Not required

Source vector

AV:A/AC:L/Au:N/C:N/I:N/A:P