Partial re-association of Philips IntelliVue MX40 monitor
Philips IntelliVue MX40 Patient Worn Monitor
At a glance
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
Source description · CVE List / CNA
System impact
CVSS summary v3.0
- Published:
- Updated:
- EPSS:
- 0.769%
- Exploited in attacks:
- Not confirmed
- Ransomware campaigns:
- Not confirmed
- Attack vector:
- Adjacent
Affected software
Vendor, product and affected version ranges.
| Vendor | Product | Affected versions |
|---|---|---|
| Vendor:Philips | IntelliVue MX40 Patient Worn Monitor | Affected versions:IntelliVue MX40 Patient Worn Monitor (WLAN only), all versions prior to Version B.06.18Show full version range for IntelliVue MX40 Patient Worn Monitor |
| Vendor:philips | intellivue mx40 firmware | Affected versions:< b.06.18Show full version range for intellivue mx40 firmware
|
1–2 of 2 · Page 1 of 1
CVSS scores and metrics2
[email protected] · NVD (Primary)Primary6.5 / 10MediumCVSS 3.0Details
- Attack complexity
- Low
- Confidentiality
- None
- Integrity
- None
- Availability
- High
- Attack vector
- Adjacent
- Privileges required
- Not required
- User interaction
- Not required
- Scope
- Unchanged
Source vector
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H[email protected] · NVD (Primary)3.3 / 10LowCVSS 2.0Details
- Attack complexity
- Low
- Confidentiality
- None
- Integrity
- None
- Availability
- Partial
- Attack vector
- Adjacent
- Authentication
- Not required
Source vector
AV:A/AC:L/Au:N/C:N/I:N/A:PCVSS measures severity; EPSS estimates exploitation probability. Lack of confirmed attacks does not mean lack of risk.
Sources and advisories6