Failure to verify VPN connection before downloading updates
Medtronic 2090 CareLink Programmer
At a glance
Medtronic 2090 CareLink Programmer uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.
Source description · CVE List / CNA
System impact
CVSS summary v3.1
- Published:
- Updated:
- EPSS:
- 1.33%
- Exploited in attacks:
- Not confirmed
- Ransomware campaigns:
- Not confirmed
- Attack vector:
- Network
Affected software
Vendor, product and affected version ranges.
| Vendor | Product | Affected versions |
|---|---|---|
| Vendor:Medtronic | 2090 CareLink Programmer | Affected versions: All versions |
| Vendor:Medtronic | 29901 Encore Programmer | Affected versions: All versions |
| Vendor:medtronic | 2090 carelink programmer firmware | Affected versions:Range not specifiedShow full version range for 2090 carelink programmer firmware
|
1–3 of 3 · Page 1 of 1
CVSS scores and metrics4
icscert · CVE List / CNA (cna)Primary7.1 / 10HighCVSS 3.1Details
- Attack complexity
- High
- Confidentiality
- High
- Integrity
- High
- Availability
- High
- Attack vector
- Network
- Privileges required
- Low
- User interaction
- Required
- Scope
- Unchanged
Source vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H[email protected] · NVD (Secondary)7.1 / 10HighCVSS 3.1Details
- Attack complexity
- High
- Confidentiality
- High
- Integrity
- High
- Availability
- High
- Attack vector
- Network
- Privileges required
- Low
- User interaction
- Required
- Scope
- Unchanged
Source vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H[email protected] · NVD (Primary)8.0 / 10HighCVSS 3.0Details
- Attack complexity
- Low
- Confidentiality
- High
- Integrity
- High
- Availability
- High
- Attack vector
- Adjacent
- Privileges required
- Low
- User interaction
- Not required
- Scope
- Unchanged
Source vector
CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H[email protected] · NVD (Primary)5.2 / 10MediumCVSS 2.0Details
- Attack complexity
- Low
- Confidentiality
- Partial
- Integrity
- Partial
- Availability
- Partial
- Attack vector
- Adjacent
- Authentication
- Single
Source vector
AV:A/AC:L/Au:S/C:P/I:P/A:PCVSS measures severity; EPSS estimates exploitation probability. Lack of confirmed attacks does not mean lack of risk.
Sources and advisories6
This product uses the NVD API but is not endorsed or certified by the NVD. CVE® records © The MITRE Corporation, reproduced under the CVE Terms of Use. Data sources and licences